Sterilization is 무균 조직이나 혈관계와 접촉하는 의료기기의 환자 안전을 보장하기 위해 생존 가능한 미생물을 제거하는 검증된 공정.
Complete Guide to Sterilization
Sterilization is a critical process for medical devices that come into contact with sterile body tissues or the vascular system. It involves validated methods to achieve a defined sterility assurance level (SAL), typically 10⁻⁶, meaning the probability of a viable microorganism being present is less than one in a million.
Common sterilization methods:
1. Ethylene Oxide (EtO) Sterilization
- Low-temperature process (37-63°C)
- Suitable for heat-sensitive materials
- Penetrates packaging and device lumens
- Regulated by ISO 11135
- Requires aeration to remove residuals
- Environmental concerns due to emissions
2. Gamma Radiation Sterilization
- Uses Cobalt-60 isotope radiation
- Room temperature process
- Excellent material penetration
- Suitable for sealed, final packages
- Regulated by ISO 11137-1, 11137-2
- Can affect polymer properties
3. E-beam (Electron Beam) Sterilization
- Uses accelerated electrons
- Faster than gamma radiation
- Lower penetration depth
- Regulated by ISO 11137-1, 11137-2
- More precise dose control
- Growing popularity for sustainability
4. Steam Sterilization (Autoclave)
- High temperature (121-134°C) with saturated steam
- Most economical method
- Limited to heat-stable materials
- Regulated by ISO 17665
- Common for reusable surgical instruments
- Short cycle times
Sterility Assurance Level (SAL):
The probability of a single viable microorganism occurring on a product after sterilization:
- SAL 10⁻⁶ - Standard for most medical devices
- SAL 10⁻³ - Low-temperature sterilization, some Class I devices
Validation requirements:
Manufacturers must demonstrate:
- Installation Qualification (IQ)
- Operational Qualification (OQ)
- Performance Qualification (PQ)
- Routine monitoring and revalidation
- Bioburden testing before sterilization
- Sterility testing after sterilization
Key standards:
- ISO 11135 - EtO sterilization
- ISO 11137 (parts 1-3) - Radiation sterilization
- ISO 17665 (parts 1-3) - Steam sterilization
- ISO 11138 - Biological indicators
- ISO 11737 - Microbiological methods
Regulatory considerations:
- Must be validated according to ISO standards
- Part of Quality Management System (ISO 13485)
- Included in Design History File (DHF)
- Critical for Class II and III sterile devices
- Requires stability testing post-sterilization
Related Terms
More Compliance & Standards
View all미국 의료기기 제조업체에 대한 현행 우수 제조 관리 기준(cGMP) 요구사항을 규정하는 FDA의 품질 시스템 규정(QSR).
품질 활동 및 결과가 계획된 약정을 준수하는지 여부와 이러한 약정이 효과적으로 구현되고 있는지 확인하기 위한 품질 관리 시스템의 체계적이고 독립적인 검사.
의료기기 제조 및 운영에서 품질 문제를 조사, 시정 및 예방하기 위한 체계적인 접근 방식.
유럽경제지역에서 판매되는 의료기기에 필수인 적합성 표시로, EU 보건, 안전 및 환경 요건 준수를 나타냄.
Need Help with Global Registration?
Pure Global provides regulatory consulting and AI-powered tools to help medical device companies navigate Global market access.