IEC 62304 is 개발, 유지보수, 리스크 관리, 구성 관리를 포함한 의료기기 소프트웨어의 수명주기 요구사항을 정의하는 국제 표준.
Complete Guide to IEC 62304
IEC 62304:2006+AMD1:2015 "Medical device software - Software life cycle processes" is the primary international standard that establishes requirements for the development, maintenance, and risk management of medical device software. It applies to both software that is part of a medical device and standalone Software as a Medical Device (SaMD).
Scope and applicability:
IEC 62304 applies to:
- Embedded software in medical devices
- Software as a Medical Device (SaMD)
- Software used to manufacture or maintain medical devices
- Off-the-shelf (OTS) software incorporated into medical devices
Software safety classification:
The standard requires manufacturers to classify software based on the severity of potential harm to patients or users:
Class A - No injury or damage possible:
- Software failure cannot result in any injury
- Lowest level of rigor required
- Minimal documentation
- Example: Administrative software with no patient impact
Class B - Non-serious injury possible:
- Software failure could result in non-serious injury
- Moderate rigor in development and testing
- Moderate documentation requirements
- Example: Diagnostic imaging viewing software
Class C - Death or serious injury possible:
- Software failure could result in death or serious injury
- Highest rigor in all lifecycle activities
- Comprehensive documentation and testing
- Example: Insulin pump control software, radiation therapy planning
Key software lifecycle processes:
1. Software development planning (5.1)
- Define software development lifecycle model (Waterfall, Agile, V-model)
- Identify deliverables and milestones
- Establish standards, methods, and tools
- Define responsibilities and resources
2. Software requirements analysis (5.2)
- Define and document software requirements
- Identify functional and performance requirements
- Define software system inputs and outputs
- Include risk control measures from ISO 14971
- Establish requirements traceability
3. Software architectural design (5.3)
- Transform requirements into architecture
- Define major software components and interfaces
- Identify software items to be developed vs. OTS
- Verify architecture implements requirements
4. Software detailed design (5.4)
- Refine architecture into detailed designs
- Design each software unit
- Specify interfaces between units
- Enable coding and testing at unit level
5. Software unit implementation and verification (5.5)
- Implement each software unit
- Verify units meet detailed design
- Establish coding standards
- Conduct unit testing
6. Software integration and integration testing (5.6)
- Integrate software units into larger items
- Verify integrated items work together
- Test interfaces and data flow
- Regression testing after integration
7. Software system testing (5.7)
- Test complete integrated software system
- Verify all software requirements are met
- Include worst-case and boundary testing
- Document test cases and results
8. Software release (5.8)
- Ensure all activities are complete
- Archive configuration items and documentation
- Create release notes and known anomalies list
- Obtain approval for release
Software maintenance process:
- Problem and modification analysis
- Modification implementation
- Maintenance review, approval, and release
- Migration to new environments
- Software retirement
Software risk management:
IEC 62304 requires integration with ISO 14971 risk management:
- Identify software hazards and hazardous situations
- Implement risk control measures in software
- Verify effectiveness of risk controls
- Document residual risks
Software configuration management:
- Configuration identification and baselines
- Change control procedures
- Version control systems
- Configuration status accounting
- Release management
Software problem resolution:
- Problem reporting system
- Problem analysis and evaluation
- Investigate impact on safety
- Implement solutions and verify effectiveness
- Trend analysis for systemic issues
Relationship to regulations:
FDA recognition:
- Recognized consensus standard
- Referenced in premarket guidance documents
- Acceptable approach for software development
- Used for 510(k), PMA, De Novo submissions
EU MDR/IVDR compliance:
- Harmonized standard supporting Annex I requirements
- Demonstrates conformity for CE marking
- Required by Notified Bodies for software-containing devices
Integration with ISO 13485:
Software development per IEC 62304 supports ISO 13485 design control requirements (Clause 7.3).
Documentation requirements:
Key documents required by IEC 62304:
- Software development plan
- Software requirements specification (SRS)
- Software architecture document
- Software detailed design document
- Software verification and validation plans and reports
- Risk management file (per ISO 14971)
- Software configuration management plan
- Problem resolution and maintenance records
Off-the-shelf (OTS) software:
Special requirements for commercial or open-source software:
- Validate OTS software for intended use
- Document OTS software identification and version
- Assess OTS software anomalies and risks
- Monitor OTS software updates and patches
Agile and modern methodologies:
IEC 62304 does not prescribe a specific development methodology:
- Compatible with Agile, DevOps, continuous integration
- Focus is on activities and deliverables, not process
- Risk-based approach allows flexibility
- Maintain traceability throughout iterations
Common compliance challenges:
- Maintaining traceability across all lifecycle phases
- Adequate testing coverage for Class C software
- Change control for rapid software updates
- Managing third-party and open-source components
- Balancing rigor with development efficiency
IEC 62304 compliance demonstrates that software development follows a systematic, risk-based approach appropriate for medical devices, providing confidence to regulators, notified bodies, and patients that the software is safe and effective.
Related Terms
More Compliance & Standards
View all미국 의료기기 제조업체에 대한 현행 우수 제조 관리 기준(cGMP) 요구사항을 규정하는 FDA의 품질 시스템 규정(QSR).
품질 활동 및 결과가 계획된 약정을 준수하는지 여부와 이러한 약정이 효과적으로 구현되고 있는지 확인하기 위한 품질 관리 시스템의 체계적이고 독립적인 검사.
의료기기 제조 및 운영에서 품질 문제를 조사, 시정 및 예방하기 위한 체계적인 접근 방식.
유럽경제지역에서 판매되는 의료기기에 필수인 적합성 표시로, EU 보건, 안전 및 환경 요건 준수를 나타냄.
Need Help with Global Registration?
Pure Global provides regulatory consulting and AI-powered tools to help medical device companies navigate Global market access.