Oil-immersed kV Tank - Vietnam Registration 230003407/PCBB-HN
Access comprehensive regulatory information for Oil-immersed kV Tank in Vietnam's medical device market through Pure Global AI's free database. This Risk Class TTBYT Loại B medical device is registered under number 230003407/PCBB-HN and manufactured by Công Ty THHH XQ Toàn Thắng.
This page provides complete technical specifications, regulatory compliance details. Pure Global AI offers free access to 300,000+ Vietnamese medical device registrations, helping global MedTech companies navigate Vietnam MOH regulations, identify competitors, and discover partnership opportunities efficiently.
Device Details
Technical Details
- This device is designed to operate as an integral part of the X-ray imaging system; Model: BT 300, but it does not have the direct function of emitting X-rays. - This device can be considered as an AC/DC step-up transformer. - This device operates by being connected through the control cabinet and the X-ray tube head. The control cabinet supplies an AC voltage to the primary winding with the appropriate selection parameters, and the device then steps up the voltage to the secondary winding through full-wave rectification, resulting in an output DC voltage in kilovolts (kV). The input and output voltage ratios are linearly proportional to each other. - It operates to supply high voltage to the X-ray tube. - This oil-immersed KV generator housing design also incorporates the hairpin transformer winding, which supplies the filament heating voltage to the X-ray tube head.
Dates and Status
Dec 26, 2023